温度湿度试验箱小型 [温度、湿度试验箱校准指标定义的研究]

何庆芳,肖 竞,曲雯洁

( 中国电器科学研究院有限公司威凯检测技术有限公,510300)

摘要:有关温度、湿度试验箱的偏差、均匀度、波动度等技术指标在JJF1101-2003 环境试验设备校准规范及JJF1270-2010 温度湿度振动综合环境试验系统校准规范都做出了相应的规定,针对这两个校准规范对偏差、均匀度、波动度的定义,使用过程中发现容易混淆,我们对这些定义做了深入的研究。

关键词:温度偏差;波动度

Definition of calibration index of temperature and humidity test chamber

He Qingfang,Xiao Jing,Qu Wenjie

(Chinese Electrical Science Research Institute Co., limited 510300 CVC detection technology)

Abstract :The deviation of the temperature and humidity test chamber,uniformity,volatility specifications in JJF1101-2003 environmental test equipment calibration specification and JJF1270-2010 temperature, humidity and vibration combined environment test system calibration specification made corresponding provisions,for the two calibration specification of deviation,uniformity,volatility is defined,using found confusing,we of these definitions do in-depth research.

Keywords :temperature deviation;fluctuation degree

1 概念的比较

我们先看看JJF1101-2003 和JJF1270-2010 偏差、均匀度、波动度的定义,见表1。

温度指示仪表示值误差检定:依据规程JJG847-2007 温度指示仪控制仪检定规程的4.1 条示值误差定义及6.1.1 条检定设备和6.3.3 条示值误差的检定,使用的校准恒温槽要求是:

工作区域最大温差:0.30℃ (-60℃~室温),0.1℃ ( 室温~ 90℃ ),0.2℃ (90℃~ 300℃ ) ;

工作区域水平温差: 0.15℃ (-60℃~室温),0.05( 室温~ 90℃ ),0.10℃ (90℃~ 300℃ ) ;

控温波动度:±0.05℃。

通常的温度试验箱的温度均匀度为2 ℃ , 波动度为±0.5℃ , 显然JJF1270-2010 的8.5 条规定来校准温度指示仪表示值误差是不符合要求的。而JJF1270-2010 的8.5 条规定的: 设备指示温度的平均值与设备工作空间全部测量点的温度测量

表1 温度偏差、均匀度、波动度的定义

(下转4 页)41

理论与算法